VEDR01 Versa DR
Implantable Pulse Generator (IPG)
VEDR01 Versa DR
Information

Device Survival Probability

VEDR01_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr at 189.0 mo          
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0          
% Including Normal Battery Depletion 99.9 99.9 99.8 99.6 99.2 98.6 97.2 92.6 81.6 67.6 54.3 43.6 37.6 34.9 34.0 33.6          
#   98637 90146 82056 74674 67967 62034 55730 47177 35507 25260 16918 10624 6616 3749 1570 181          
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PWH, NWH
Mass 27.1 g
Volume 12.1 cc
Waveform N/A
X-Ray ID PWH, PWB
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.4 | 9.1* 8.2 | 10.3*
Nominal 3.5 V (A, RV) 6.0 | 7.2* 7.4 | 9.0*
High 5.0 V (A, RV) 4.5 | 5.1* 6.0 | 7.0*
Footnote

*The second longevity numbers provided apply to the N-series Versa VEDR01 product. The N-series products have increased longevity because of improved electronic circuitry. Click on "Other Articles" for more information on the N-series longevity. Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 %Hide this content

Information

Distribution Data

US Market Release2006-07-17
CE Approval Date2005-09-20
Registered USA Implants118957
Estimated Active USA Implants23370
Normal Battery Depletions15041
US Market Release2006-07-17
CE Approval Date2005-09-20
Estimated WW Distribution178057
Normal Battery Depletions15070
Information

Malfunctions (USA)

Therapy Function Not Compromised 13
Electrical Component 9
Electrical Interconnect 2
Possible Early Battery Depletion 2
Therapy Function Compromised 15
Electrical Component 11
Other 4
Information

Malfunctions (WW)

Therapy Function Not Compromised 20
Electrical Component 16
Electrical Interconnect 2
Possible Early Battery Depletion 2
Therapy Function Compromised 20
Electrical Component 15
Electrical Interconnect 1
Other 4
Information

EOS Indication

Prolonged Service period is 3 months after RRT/ERI assuming 100% pacing at the listed parameter settings. Reprogramming of pacing parameters may reduce the duration of the prolonged service period.

Information

Recommended Replacement (RRT)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of May 24, 2025

www.medtronic.com/productperformance 1051050