Customer Communications For This Model
- There are no communications for this model.
US Acute Lead Observations
| Failure to Capture |
1 |
| Failure to Sense |
7 |
| Impedance Out of Range |
31 |
| Lead Dislodgement |
20 |
| Oversensing |
66 |
Product Characteristics
| Lead Type |
Defibrillation |
| Type and/or Fixation |
Shaped passive fixation |
| Lead Function |
Pacing/Sensing and Defibrillation |
| Lead Placement |
Extravascular |
| Lead Tip Location |
Anterior mediastinum |
| Pace/Sense Polarity |
True Bipolar/Two Coils |
More
| Serial Number Prefix |
EVL |
| Lead Body Diameter |
2.9 mm (8.7 Fr) |
| Insulation |
Polyurethane |
| Pace/Sense Connector Style |
EV4-LLHH |
| High Voltage Connector |
EV4-LLHH |
| Steroid |
None |
Hide this content