5086MRI CapsureFix Novus MRI
Pacing
5086MRI CapsureFix Novus MRI
VENTRICULAR PLACEMENT
Information

Device Survival Probability

5086MRI_VEN_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr at 162.0 mo              
% Survival 99.7 99.7 99.6 99.6 99.6 99.3 98.3 98.0 98.0 98.0 98.0 97.6 97.6 97.6              
#   2525 2182 1850 1426 734 422 374 331 300 278 220 160 83 51              

Product Surveillance Registry Results

Number of Leads Enrolled in Study 3072
Cumulative Months of Follow-Up 147318
Number of Leads Active in Study 1217
Information 20

Qualifying Complications

Conductor Fracture 3
Failure to Capture 8
Failure to Sense 2
Impedance Out of Range 2
Lead Dislodgement 3
Oversensing 1
Other 1
ATRIAL PLACEMENT
Information

Device Survival Probability

5086MRI_ATR_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr at 162.0 mo              
% Survival 99.8 99.6 99.6 99.4 99.3 98.9 98.4 98.4 98.4 98.1 98.1 98.1 98.1 98.1              
#   2528 2202 1878 1461 766 453 398 354 322 297 239 173 86 50              

Product Surveillance Registry Results

Number of Leads Enrolled in Study 3142
Cumulative Months of Follow-Up 150575
Number of Leads Active in Study 1233
Information 21

Qualifying Complications

Conductor Fracture 3
Failure to Capture 3
Lead Dislodgement 12
Oversensing 2
Other 1
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 213
Conductor Fracture 4
Extra Cardiac Stimulation 18
Failure to Capture 145
Failure to Sense 29
Impedance Out of Range 10
Insulation Breach 2
Lead Dislodgement 312
Oversensing 32
Information

Distribution Data

US Market Release2011-02-08
CE Approval Date2009-01-21
Registered USA Implants207839
Estimated Active USA Implants123757
US Market Release2011-02-08
CE Approval Date2009-01-21
Estimated WW Distribution398389
Information

USA Returned Product Analysis

Conductor Fracture 125
Insulation Breach 235
Crimp/Weld/Bond 0
Other 12
Information

WW Returned Product Analysis

Conductor Fracture 149
Insulation Breach 258
Crimp/Weld/Bond 0
Other 17
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium or Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LFP
Lead Body Diameter 6.9 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of June 8, 2026

www.medtronic.com/productperformance 22521