5086MRI
CapsureFix Novus MRI
VENTRICULAR PLACEMENT
Device Survival Probability
5086MRI_VEN_SURV
Loading...
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | at 162.0 mo | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.7 | 99.7 | 99.6 | 99.6 | 99.6 | 99.3 | 98.3 | 98.0 | 98.0 | 98.0 | 98.0 | 97.6 | 97.6 | 97.6 | ||||||||
| # | 2525 | 2182 | 1850 | 1426 | 734 | 422 | 374 | 331 | 300 | 278 | 220 | 160 | 83 | 51 |
Product Surveillance Registry Results
| Number of Leads Enrolled in Study | 3072 |
|---|---|
| Cumulative Months of Follow-Up | 147318 |
| Number of Leads Active in Study | 1217 |
ATRIAL PLACEMENT
Device Survival Probability
5086MRI_ATR_SURV
Loading...
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | at 162.0 mo | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.8 | 99.6 | 99.6 | 99.4 | 99.3 | 98.9 | 98.4 | 98.4 | 98.4 | 98.1 | 98.1 | 98.1 | 98.1 | 98.1 | ||||||||
| # | 2528 | 2202 | 1878 | 1461 | 766 | 453 | 398 | 354 | 322 | 297 | 239 | 173 | 86 | 50 |
Product Surveillance Registry Results
| Number of Leads Enrolled in Study | 3142 |
|---|---|
| Cumulative Months of Follow-Up | 150575 |
| Number of Leads Active in Study | 1233 |
US Acute Lead Observations
| Cardiac Perforation | 213 |
|---|---|
| Conductor Fracture | 4 |
| Extra Cardiac Stimulation | 18 |
| Failure to Capture | 145 |
| Failure to Sense | 29 |
| Impedance Out of Range | 10 |
| Insulation Breach | 2 |
| Lead Dislodgement | 312 |
| Oversensing | 32 |
Product Characteristics
| Lead Type | Pacing |
|---|---|
| Type and/or Fixation | Active Screw In |
| Lead Function | Pacing/Sensing |
| Lead Placement | Transvenous |
| Lead Tip Location | Atrium or Right Ventricle |
| Pace/Sense Polarity | Bipolar |
| Serial Number Prefix | LFP |
|---|---|
| Lead Body Diameter | 6.9 FR |
| Insulation | Silicone |
| Pace/Sense Connector Style | IS-1 BI |
| High Voltage Connector | n/a |
| Steroid | Yes |
Data as of June 8, 2026