6721 Epicardial Patch
Defibrillation
ALL
Device Survival Probability
6721_6921_SURV
Loading...
| |
|
1 yr |
2 yr |
3 yr |
4 yr |
5 yr |
6 yr |
7 yr |
8 yr |
9 yr |
|
| % |
 |
96.6 |
94.6 |
92.1 |
91.0 |
89.3 |
84.7 |
83.3 |
83.3 |
83.3 |
|
| # |
|
347 |
319 |
274 |
221 |
190 |
136 |
102 |
66 |
57 |
|
Product Surveillance Registry Results
| Number of Leads Enrolled in Study |
417 |
| Cumulative Months of Follow-Up |
24242 |
| Number of Leads Active in Study |
5 |
51
Qualifying Complications
| Conductor Fracture |
21 |
| Failure to Capture |
8 |
| Impedance Out of Range |
4 |
| Insulation (not further defined) |
2 |
| Other |
16 |
Customer Communications For This Model
- There are no communications for this model.
US Acute Lead Observations
| Cardiac Perforation |
1 |
| Conductor Fracture |
2 |
| Failure to Capture |
4 |
| Failure to Sense |
2 |
| Impedance Out of Range |
24 |
| Oversensing |
1 |
Product Characteristics
| Lead Type |
Defibrillation |
| Type and/or Fixation |
Suture |
| Lead Function |
Defibrillation Electrode |
| Lead Placement |
Epi Patch |
| Lead Tip Location |
Epicardial |
| Pace/Sense Polarity |
n/a |
More
| Serial Number Prefix |
TBB, TBG, TBH |
| Lead Body Diameter |
S, M, L |
| Insulation |
Silicone |
| Pace/Sense Connector Style |
n/a |
| High Voltage Connector |
DF-1 |
| Steroid |
None |