Device Survival Probability
6721_6921_SURV
Loading...
|
|
1 yr |
2 yr |
3 yr |
4 yr |
5 yr |
6 yr |
7 yr |
8 yr |
9 yr |
|
% |
 |
96.6 |
94.6 |
92.1 |
91.0 |
89.3 |
84.7 |
83.3 |
83.3 |
83.3 |
|
# |
|
347 |
319 |
274 |
221 |
190 |
136 |
102 |
66 |
57 |
|
Product Surveillance Registry Results
Number of Leads Enrolled in Study |
417 |
Cumulative Months of Follow-Up |
24242 |
Number of Leads Active in Study |
5 |
51
Qualifying Complications
Conductor Fracture |
21 |
Failure to Capture |
8 |
Impedance Out of Range |
4 |
Insulation (not further defined) |
2 |
Other |
16 |
Customer Communications For This Model
- There are no communications for this model.
US Acute Lead Observations
Cardiac Perforation |
1 |
Conductor Fracture |
2 |
Failure to Capture |
4 |
Failure to Sense |
2 |
Impedance Out of Range |
24 |
Oversensing |
1 |
Product Characteristics
Lead Type |
Defibrillation |
Type and/or Fixation |
Suture |
Lead Function |
Defibrillation Electrode |
Lead Placement |
Epi Patch |
Lead Tip Location |
Epicardial |
Pace/Sense Polarity |
n/a |
More
Serial Number Prefix |
TBB, TBG, TBH |
Lead Body Diameter |
S, M, L |
Insulation |
Silicone |
Pace/Sense Connector Style |
n/a |
High Voltage Connector |
DF-1 |
Steroid |
None |