4074 CapSure Sense
Pacing
4074 CapSure Sense
VENTRICULAR PLACEMENT
Information

Device Survival Probability

4074_VEN_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr 17 yr 18 yr      
% Survival 99.5 99.3 99.2 99.2 99.2 99.2 99.2 99.2 99.2 98.8 98.8 97.6 97.6 96.9 96.9 96.9 96.9 96.9      
#   1031 880 742 638 499 408 344 300 261 223 193 161 130 123 111 92 73 50      

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1192
Cumulative Months of Follow-Up 82469
Number of Leads Active in Study 110
Information 12

Qualifying Complications

Conductor Fracture 1
Failure to Capture 4
Impedance Out of Range 2
Insulation (not further defined) 2
Lead Dislodgement 2
Other 1
ATRIAL PLACEMENT
Information

Device Survival Probability

4074_ATR_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr 17 yr at 210.0 mo      
% Survival 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1      
#   214 205 198 183 167 158 148 136 126 117 110 107 99 96 88 77 68 61      

Product Surveillance Registry Results

Number of Leads Enrolled in Study 227
Cumulative Months of Follow-Up 30107
Number of Leads Active in Study 45
Information 2

Qualifying Complications

Failure to Sense 1
Lead Dislodgement 1
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 38
Conductor Fracture 2
Extra Cardiac Stimulation 4
Failure to Capture 200
Failure to Sense 19
Impedance Out of Range 16
Lead Dislodgement 215
Oversensing 9
Information

Distribution Data

US Market Release2002-06-23
CE Approval Date2002-02-01
Registered USA Implants159708
Estimated Active USA Implants77699
US Market Release2002-06-23
CE Approval Date2002-02-01
Estimated WW Distribution1124929
Information

USA Returned Product Analysis

Conductor Fracture 16
Insulation Breach 64
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 23
Insulation Breach 69
Crimp/Weld/Bond 0
Other 5
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix BBD
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of May 24, 2025

www.medtronic.com/productperformance 1051050