4074
CapSure Sense
VENTRICULAR PLACEMENT

Device Survival Probability
4074_VEN_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | 15 yr | 16 yr | 17 yr | 18 yr | |||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.5 | 99.3 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 99.2 | 98.8 | 98.8 | 97.6 | 97.6 | 96.9 | 96.9 | 96.9 | 96.9 | 96.9 | |||
# | 1031 | 880 | 742 | 638 | 499 | 408 | 344 | 300 | 261 | 223 | 193 | 161 | 130 | 123 | 111 | 92 | 73 | 50 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 1192 |
---|---|
Cumulative Months of Follow-Up | 82469 |
Number of Leads Active in Study | 110 |
ATRIAL PLACEMENT

Device Survival Probability
4074_ATR_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | 12 yr | 13 yr | 14 yr | 15 yr | 16 yr | 17 yr | at 210.0 mo | |||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | 99.1 | |||
# | 214 | 205 | 198 | 183 | 167 | 158 | 148 | 136 | 126 | 117 | 110 | 107 | 99 | 96 | 88 | 77 | 68 | 61 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 227 |
---|---|
Cumulative Months of Follow-Up | 30107 |
Number of Leads Active in Study | 45 |

US Acute Lead Observations
Cardiac Perforation | 38 |
---|---|
Conductor Fracture | 2 |
Extra Cardiac Stimulation | 4 |
Failure to Capture | 200 |
Failure to Sense | 19 |
Impedance Out of Range | 16 |
Lead Dislodgement | 215 |
Oversensing | 9 |

Product Characteristics
Lead Type | Pacing |
---|---|
Type and/or Fixation | Tines |
Lead Function | Pacing/Sensing |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | Bipolar |
Serial Number Prefix | BBD |
---|---|
Lead Body Diameter | 5.3 FR |
Insulation | Polyurethane (outer) Silicone (inner) |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | n/a |
Steroid | Yes |
Data as of May 24, 2025