4296 Attain Ability Plus
Left-Heart Pacing
4296 Attain Ability Plus
ALL
Information

Device Survival Probability

4296_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr                  
% Survival 98.7 97.9 97.7 97.4 96.8 96.4 96.4 96.1 96.1 95.8 95.8 95.8                  
#   1163 941 776 660 556 474 409 330 284 225 138 52                  

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1473
Cumulative Months of Follow-Up 80295
Number of Leads Active in Study 168
Information 38

Qualifying Complications

Conductor Fracture 1
Extra Cardiac Stimulation 12
Failure to Capture 9
Lead Dislodgement 14
Other 2
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 2
Conductor Fracture 1
Extra Cardiac Stimulation 65
Failure to Capture 39
Impedance Out of Range 11
Insulation Breach 4
Lead Dislodgement 120
Information

Distribution Data

US Market Release2011-04-01
CE Approval Date2009-12-18
Registered USA Implants35324
Estimated Active USA Implants17117
US Market Release2011-04-01
CE Approval Date2009-12-18
Estimated WW Distribution124279
Information

USA Returned Product Analysis

Conductor Fracture 4
Insulation Breach 0
Crimp/Weld/Bond 2
Other 4
Information

WW Returned Product Analysis

Conductor Fracture 10
Insulation Breach 0
Crimp/Weld/Bond 2
Other 4
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Dual Electrodes
More
Serial Number Prefix RRA
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of May 24, 2025

www.medtronic.com/productperformance 1051050