4296 Attain Ability Plus
Left-Heart Pacing
4296 Attain Ability Plus
ALL
Information

Device Survival Probability

4296_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr at 150.0 mo                
% Survival 98.6 97.8 97.6 97.3 96.7 96.3 96.3 96.1 96.1 95.7 95.7 95.7 95.7                
#   1164 943 777 662 558 477 411 335 285 234 185 91 56                

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1473
Cumulative Months of Follow-Up 82028
Number of Leads Active in Study 149
Information 37

Qualifying Complications

Conductor Fracture 1
Extra Cardiac Stimulation 11
Failure to Capture 9
Lead Dislodgement 14
Other 2
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 2
Conductor Fracture 1
Extra Cardiac Stimulation 65
Failure to Capture 41
Impedance Out of Range 11
Insulation Breach 4
Lead Dislodgement 120
Information

Distribution Data

US Market Release2011-04-01
CE Approval Date2009-12-18
Registered USA Implants35450
Estimated Active USA Implants17033
US Market Release2011-04-01
CE Approval Date2009-12-18
Estimated WW Distribution126826
Information

USA Returned Product Analysis

Conductor Fracture 4
Insulation Breach 0
Crimp/Weld/Bond 2
Other 4
Information

WW Returned Product Analysis

Conductor Fracture 10
Insulation Breach 0
Crimp/Weld/Bond 2
Other 4
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Dual Electrodes
More
Serial Number Prefix RRA
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) SI-polyimide (SI-PI)(inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of June 8, 2026

www.medtronic.com/productperformance 22521