4965 CapSure Epi
Epi/Myocardial Pacing
4965 CapSure Epi
ALL
Information

Device Survival Probability

4965_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr            
% Survival 97.9 95.1 94.0 85.6            
#   118 100 82 60            

Product Surveillance Registry Results

Number of Leads Enrolled in Study 234
Cumulative Months of Follow-Up 7543
Number of Leads Active in Study 3
Information 15

Qualifying Complications

Conductor Fracture 10
Failure to Capture 3
Failure to Sense 1
Insulation (not further defined) 1
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 1
Conductor Fracture 1
Failure to Capture 13
Failure to Sense 8
Impedance Out of Range 21
Oversensing 4
Unspecified Clinical Failure 3
Information

Distribution Data

US Market Release1996-09-06
CE Approval Date1993-01-01
Registered USA Implants24442
Estimated Active USA Implants6654
US Market Release1996-09-06
CE Approval Date1993-01-01
Estimated WW Distribution65177
Information

USA Returned Product Analysis

Conductor Fracture 307
Insulation Breach 66
Crimp/Weld/Bond 1
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 366
Insulation Breach 76
Crimp/Weld/Bond 1
Other 0
Information

Product Characteristics

Lead Type Epi/Myocardial
Type and/or Fixation Suture
Lead Function Pacing/Sensing
Lead Placement Myocardial
Lead Tip Location Atrium or Right Ventricle
Pace/Sense Polarity Unipolar
More
Serial Number Prefix LBT
Lead Body Diameter 4.5 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of June 8, 2026