DVFB1D4 Visia MRI AF
Implantable Cardioverter Defibrillator (ICD)
DVFB1D4 Visia MRI AF
Information

Device Survival Probability

VISIA_AF_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 107.0 mo  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 99.9 99.9 99.8 99.8 99.7 99.6  
% Including Normal Battery Depletion 100.0 99.9 99.9 99.8 99.7 99.6 99.4 99.0 98.8  
#   76507 71635 64618 55414 47096 35528 22819 10781 578  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 36 J
More
Connector Style DF4 VR Style
Serial Number Prefix PKX
Mass 77 g
Volume 33 cc
Waveform Biphasic
X-Ray ID PFZ
Hide this content
DF4 VR Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.9 5.3 5.7 5.8
Quarterly 7.8 8.8 9.7 10.1
Semi-annual 8.5 9.7 10.7 11.3
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release2016-01-19
CE Approval Date
Registered USA Implants57296
Estimated Active USA Implants45213
Normal Battery Depletions53
US Market Release2016-01-19
CE Approval Date
Estimated WW Distribution76929
Normal Battery Depletions53
Information

Malfunctions (USA)

Therapy Function Not Compromised 62
Battery 50
Electrical Component 11
Other 1
Therapy Function Compromised 34
Battery 16
Device-Related Current Pathway 15
Electrical Component 3
Information

Malfunctions (WW)

Therapy Function Not Compromised 62
Battery 50
Electrical Component 11
Other 1
Therapy Function Compromised 34
Battery 16
Device-Related Current Pathway 15
Electrical Component 3
Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than 2.73V on 3 consecutive daily automatic measurements.

Information

Charge Time

BLACKWELL_VR_CHRG
Loading...

Data as of December 1, 2025

www.medtronic.com/productperformance 22021