4798 Attain Stability Quad
Left-Heart Pacing
4798 Attain Stability Quad
ALL
Information

Device Survival Probability

4798_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 78.0 mo      
% Survival 99.7 99.4 99.4 99.2 99.2 99.2 99.2      
#   1256 825 492 305 142 74 51      

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1535
Cumulative Months of Follow-Up 45498
Number of Leads Active in Study 856
Information 9

Qualifying Complications

Conductor Fracture 1
Extra Cardiac Stimulation 1
Failure to Capture 4
Lead Dislodgement 3
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 10
Conductor Fracture 2
Extra Cardiac Stimulation 134
Failure to Capture 171
Impedance Out of Range 59
Lead Dislodgement 148
Oversensing 1
Information

Distribution Data

US Market Release2019-06-03
CE Approval Date2017-04-24
Registered USA Implants77286
Estimated Active USA Implants71270
US Market Release2019-06-03
CE Approval Date2017-04-24
Estimated WW Distribution200442
Information

USA Returned Product Analysis

Conductor Fracture 1
Insulation Breach 0
Crimp/Weld/Bond 0
Other 18
Information

WW Returned Product Analysis

Conductor Fracture 1
Insulation Breach 0
Crimp/Weld/Bond 0
Other 30
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Non-electrically Active Side Fixation
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Quadripolar
More
Serial Number Prefix QFX
Lead Body Diameter 4.4 FR
Insulation Polyurethane (outer) SI-polyimide (SI-PI)(inner)
Pace/Sense Connector Style IS4-LLLL
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of June 8, 2026

www.medtronic.com/productperformance 22521