4798 Attain Stability Quad
Left-Heart Pacing
4798 Attain Stability Quad
ALL
Information

Device Survival Probability

4798_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr at 66.0 mo        
% Survival 99.7 99.5 99.5 99.1 99.1 99.1        
#   1055 582 333 168 89 61        

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1505
Cumulative Months of Follow-Up 34824
Number of Leads Active in Study 908
Information 9

Qualifying Complications

Conductor Fracture 1
Extra Cardiac Stimulation 2
Failure to Capture 4
Lead Dislodgement 2
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 10
Conductor Fracture 2
Extra Cardiac Stimulation 122
Failure to Capture 143
Impedance Out of Range 47
Lead Dislodgement 124
Oversensing 1
Information

Distribution Data

US Market Release2019-06-03
CE Approval Date2017-04-24
Registered USA Implants63583
Estimated Active USA Implants59000
US Market Release2019-06-03
CE Approval Date2017-04-24
Estimated WW Distribution164966
Information

USA Returned Product Analysis

Conductor Fracture 1
Insulation Breach 0
Crimp/Weld/Bond 0
Other 15
Information

WW Returned Product Analysis

Conductor Fracture 1
Insulation Breach 0
Crimp/Weld/Bond 0
Other 22
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Non-electrically Active Side Fixation
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Quadripolar
More
Serial Number Prefix QFX
Lead Body Diameter 4.4 FR
Insulation Polyurethane (outer) SI-polyimide (SI-PI)(inner)
Pace/Sense Connector Style IS4-LLLL
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of May 24, 2025