6947 Sprint Quattro Secure
Defibrillation
6947 Sprint Quattro Secure
ALL
Information

Device Survival Probability

6947_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr 17 yr 18 yr 19 yr 20 yr  
% Survival 99.5 99.3 99.0 98.7 98.2 97.9 97.5 97.1 96.7 96.2 95.9 95.5 95.1 94.7 94.4 92.9 92.9 91.4 89.5 86.9  
#   3308 2909 2552 2262 2032 1788 1547 1387 1244 1091 926 742 595 437 287 187 127 111 85 59  

Product Surveillance Registry Results

Number of Leads Enrolled in Study 4600
Cumulative Months of Follow-Up 305785
Number of Leads Active in Study 419
Information 105

Qualifying Complications

Cardiac Perforation 1
Conductor Fracture 42
Failure to Capture 8
Failure to Sense 2
Impedance Out of Range 13
Insulation (not further defined) 6
Lead Dislodgement 5
Oversensing 21
Other 4
Unspecified Clinical Failure 3
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 29
Conductor Fracture 26
Extra Cardiac Stimulation 2
Failure to Capture 83
Failure to Sense 36
Impedance Out of Range 61
Insulation Breach 4
Lead Dislodgement 124
Oversensing 141
Unspecified Clinical Failure 20
Information

Distribution Data

US Market Release2001-11-12
CE Approval Date2001-10-04
Registered USA Implants375772
Estimated Active USA Implants124006
US Market Release2001-11-12
CE Approval Date2001-10-04
Estimated WW Distribution618492
Information

USA Returned Product Analysis

Conductor Fracture 1441
Insulation Breach 104
Crimp/Weld/Bond 4
Other 198
Information

WW Returned Product Analysis

Conductor Fracture 1776
Insulation Breach 136
Crimp/Weld/Bond 4
Other 240
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar/Two Coils
More
Serial Number Prefix TDG
Lead Body Diameter 8.6 FR
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of May 24, 2025

www.medtronic.com/productperformance 1051050