4298 Attain Performa
Left-Heart Pacing
4298 Attain Performa
ALL
Information

Device Survival Probability

4298_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr at 126.0 mo                    
% Survival 99.1 98.7 98.5 98.5 98.2 98.1 98.1 98.1 98.1 98.1 98.1                    
#   1910 1630 1394 1176 964 766 557 404 248 105 52                    

Product Surveillance Registry Results

Number of Leads Enrolled in Study 2284
Cumulative Months of Follow-Up 122595
Number of Leads Active in Study 462
Information 32

Qualifying Complications

Extra Cardiac Stimulation 5
Failure to Capture 6
Impedance Out of Range 1
Lead Dislodgement 16
Other 4
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 8
Conductor Fracture 1
Extra Cardiac Stimulation 253
Failure to Capture 206
Failure to Sense 1
Impedance Out of Range 57
Lead Dislodgement 284
Information

Distribution Data

US Market Release2014-08-01
CE Approval Date2013-01-01
Registered USA Implants133379
Estimated Active USA Implants102520
US Market Release2014-08-01
CE Approval Date2013-01-01
Estimated WW Distribution299961
Information

USA Returned Product Analysis

Conductor Fracture 9
Insulation Breach 0
Crimp/Weld/Bond 0
Other 27
Information

WW Returned Product Analysis

Conductor Fracture 12
Insulation Breach 0
Crimp/Weld/Bond 0
Other 31
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Quadripolar
More
Serial Number Prefix QUA
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) SI-polyimide (SI-PI)(inner)
Pace/Sense Connector Style IS-4
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of June 8, 2026