5071 Screw-in
Epi/Myocardial Pacing
5071 Screw-in
ALL
Information

Device Survival Probability

5071_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr at 90.0 mo    
% Survival 95.0 91.9 90.3 90.3 89.0 87.4 86.4 85.1    
#   237 192 162 145 123 98 79 64    

Product Surveillance Registry Results

Number of Leads Enrolled in Study 475
Cumulative Months of Follow-Up 18155
Number of Leads Active in Study 44
Information 38

Qualifying Complications

Conductor Fracture 5
Extra Cardiac Stimulation 1
Failure to Capture 21
Failure to Sense 2
Impedance Out of Range 3
Lead Dislodgement 3
Oversensing 2
Other 1
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 3
Extra Cardiac Stimulation 6
Failure to Capture 126
Failure to Sense 4
Impedance Out of Range 15
Lead Dislodgement 4
Oversensing 2
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release1992-12-03
CE Approval Date1993-01-01
Registered USA Implants58041
Estimated Active USA Implants12595
US Market Release1992-12-03
CE Approval Date1993-01-01
Estimated WW Distribution117581
Information

USA Returned Product Analysis

Conductor Fracture 35
Insulation Breach 2
Crimp/Weld/Bond 0
Other 1
Information

WW Returned Product Analysis

Conductor Fracture 36
Insulation Breach 3
Crimp/Weld/Bond 1
Other 5
Information

Product Characteristics

Lead Type Epi/Myocardial
Type and/or Fixation Fixed Screw
Lead Function Pacing/Sensing
Lead Placement Myocardial
Lead Tip Location Right Ventricle
Pace/Sense Polarity Unipolar
More
Serial Number Prefix LAQ
Lead Body Diameter 6.6 FR
Insulation Silicon
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid None
Hide this content

Data as of May 24, 2025

www.medtronic.com/productperformance 1051050