4194 Attain OTW
Left-Heart Pacing
4194 Attain OTW
ALL
Information

Device Survival Probability

4194_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr at 186.0 mo          
% Survival 98.6 97.4 96.7 96.1 95.6 94.3 94.1 93.5 93.5 93.3 93.0 93.0 93.0 91.6 91.6 91.6          
#   1238 1046 897 769 697 616 506 430 369 322 264 192 154 100 69 63          

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1654
Cumulative Months of Follow-Up 101664
Number of Leads Active in Study 119
Information 68

Qualifying Complications

Conductor Fracture 2
Extra Cardiac Stimulation 11
Failure to Capture 22
Insulation (ESC) 1
Insulation (not further defined) 2
Lead Dislodgement 30
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 2
Conductor Fracture 3
Extra Cardiac Stimulation 49
Failure to Capture 43
Impedance Out of Range 9
Lead Dislodgement 153
Oversensing 2
Unspecified Clinical Failure 4
Information

Distribution Data

US Market Release2004-08-24
CE Approval Date2003-07-14
Registered USA Implants114259
Estimated Active USA Implants28267
US Market Release2004-08-24
CE Approval Date2003-07-14
Estimated WW Distribution244151
Information

USA Returned Product Analysis

Conductor Fracture 48
Insulation Breach 167
Crimp/Weld/Bond 0
Other 2
Information

WW Returned Product Analysis

Conductor Fracture 54
Insulation Breach 173
Crimp/Weld/Bond 0
Other 4
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LFG
Lead Body Diameter 6.0 FR
Insulation Polyurethane (55D) outer; silicone inner
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of May 24, 2025

www.medtronic.com/productperformance 1051050