4194 Attain OTW
Left-Heart Pacing
4194 Attain OTW
ALL
Information

Device Survival Probability

4194_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr          
% Survival 98.6 97.4 96.7 96.1 95.6 94.3 94.1 93.5 93.5 93.3 92.7 91.9 91.9 90.7 90.7 90.7          
#   1238 1046 898 769 698 617 507 432 371 324 275 209 162 121 78 58          

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1654
Cumulative Months of Follow-Up 102970
Number of Leads Active in Study 97
Information 70

Qualifying Complications

Conductor Fracture 3
Extra Cardiac Stimulation 11
Failure to Capture 23
Insulation (ESC) 1
Insulation (not further defined) 2
Lead Dislodgement 30
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 2
Conductor Fracture 3
Extra Cardiac Stimulation 49
Failure to Capture 43
Impedance Out of Range 9
Lead Dislodgement 154
Oversensing 2
Unspecified Clinical Failure 4
Information

Distribution Data

US Market Release2004-08-24
CE Approval Date2003-07-14
Registered USA Implants114262
Estimated Active USA Implants27975
US Market Release2004-08-24
CE Approval Date2003-07-14
Estimated WW Distribution244152
Information

USA Returned Product Analysis

Conductor Fracture 49
Insulation Breach 169
Crimp/Weld/Bond 0
Other 2
Information

WW Returned Product Analysis

Conductor Fracture 55
Insulation Breach 175
Crimp/Weld/Bond 0
Other 4
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Double Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Left Ventricular Cardiac Vein
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LFG
Lead Body Diameter 6.0 FR
Insulation Polyurethane (55D) outer; silicone inner
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of June 8, 2026