DVEA3E4 Aurora EV-ICD
Implantable Cardioverter Defibrillator (ICD)
DVEA3E4 Aurora EV-ICD
Information

Device Survival Probability

AURORA_EV_SURV
Loading...










    1 yr at 15.0 mo                
% Excluding Normal Battery Depletion 100.0 100.0                
% Including Normal Battery Depletion 100.0 100.0                
#   286 114                
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVI
Max. Delivered Energy 40 J
More
Serial Number Prefix EVX
Mass 77 g
Volume 33 cc
Waveform Biphasic
X-Ray ID REX
Hide this content
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly N/A* N/A* N/A* N/A*
Quarterly N/A* N/A* N/A* N/A*
Semi-annual N/A* N/A* N/A* N/A*
Footnote

*The service life projection for the device is 11.7 years. This projection is an estimate of the average expected service life based on the following assumptions: Pacing at 0%; 2 high-voltage therapies per year; Pre-arrhythmia EGM storage programmed to On for 6-months, over life of device; Wireless telemetry: 3 hours of telemetry enabled at implant, 30 min of telemetry enabled for quarterly scheduled CareLink Monitor remote sessions (if available), 1 hour of in-office wireless telemetry enabled annually; Shelf storage life of 5 months, before implant.Hide this content

Information

Distribution Data

US Market Release2023-10-20
CE Approval Date2023-02-17
Registered USA Implants1661
Estimated Active USA Implants1614
Normal Battery Depletions
US Market Release2023-10-20
CE Approval Date2023-02-17
Estimated WW Distribution4223
Normal Battery Depletions
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised
Therapy Function Compromised
Information

EOS Indicator

From the point that the RRT is set, the device will operate for approximately three months for typical device configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than 2.73 V on 3 consecutive daily automatic measurements.

Information

Charge Time

DVEA3E4_DVEX2E4_DVEX3E4_CHRG
Loading...

Data as of May 24, 2025

www.medtronic.com/productperformance 1051050