4574 CapSure Sense
Pacing
4574 CapSure Sense
ALL
Information

Device Survival Probability

4574_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr                
% Survival 99.3 99.3 99.2 99.0 98.6 98.4 98.4 98.4 98.4 98.4 98.4 98.4 98.4                
#   1437 1140 919 716 562 474 399 338 271 201 146 102 61                

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1782
Cumulative Months of Follow-Up 92195
Number of Leads Active in Study 508
Information 17

Qualifying Complications

Conductor Fracture 2
Failure to Capture 8
Lead Dislodgement 7
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 6
Conductor Fracture 2
Extra Cardiac Stimulation 1
Failure to Capture 228
Failure to Sense 103
Impedance Out of Range 14
Lead Dislodgement 313
Oversensing 21
Unspecified Clinical Failure 4
Information

Distribution Data

US Market Release2002-06-23
CE Approval Date2002-02-01
Registered USA Implants131655
Estimated Active USA Implants77913
US Market Release2002-06-23
CE Approval Date2002-02-01
Estimated WW Distribution1004991
Information

USA Returned Product Analysis

Conductor Fracture 14
Insulation Breach 26
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 20
Insulation Breach 29
Crimp/Weld/Bond 0
Other 2
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation J-shape, tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium
Pace/Sense Polarity Bipolar
More
Serial Number Prefix BBE
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of December 1, 2025

www.medtronic.com/productperformance 22021