6947 Sprint Quattro Secure
Defibrillation
6947 Sprint Quattro Secure
ALL
Information

Device Survival Probability

6947_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr 17 yr 18 yr 19 yr 20 yr  
% Survival 99.5 99.3 99.0 98.7 98.2 97.9 97.5 97.1 96.7 96.1 95.8 95.5 95.1 94.7 94.5 93.3 93.3 92.0 90.2 87.9  
#   3310 2914 2557 2263 2032 1794 1548 1391 1247 1098 938 768 637 493 363 244 156 116 92 66  

Product Surveillance Registry Results

Number of Leads Enrolled in Study 4600
Cumulative Months of Follow-Up 310463
Number of Leads Active in Study 322
Information 106

Qualifying Complications

Cardiac Perforation 1
Conductor Fracture 43
Failure to Capture 8
Failure to Sense 2
Impedance Out of Range 14
Insulation (not further defined) 6
Lead Dislodgement 5
Oversensing 20
Other 4
Unspecified Clinical Failure 3
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 29
Conductor Fracture 27
Extra Cardiac Stimulation 2
Failure to Capture 83
Failure to Sense 36
Impedance Out of Range 61
Insulation Breach 4
Lead Dislodgement 125
Oversensing 142
Unspecified Clinical Failure 20
Information

Distribution Data

US Market Release2001-11-12
CE Approval Date2001-10-04
Registered USA Implants375910
Estimated Active USA Implants122828
US Market Release2001-11-12
CE Approval Date2001-10-04
Estimated WW Distribution619719
Information

USA Returned Product Analysis

Conductor Fracture 1467
Insulation Breach 105
Crimp/Weld/Bond 4
Other 199
Information

WW Returned Product Analysis

Conductor Fracture 1806
Insulation Breach 137
Crimp/Weld/Bond 4
Other 241
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar/Two Coils
More
Serial Number Prefix TDG
Lead Body Diameter 8.6 FR
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of June 8, 2026

www.medtronic.com/productperformance 22521