6947 Sprint Quattro Secure
Defibrillation
6947 Sprint Quattro Secure
ALL
Information

Device Survival Probability

6947_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr 17 yr 18 yr 19 yr 20 yr  
% Survival 99.5 99.3 99.0 98.7 98.2 97.9 97.5 97.1 96.7 96.2 95.9 95.5 95.1 94.7 94.5 93.2 93.2 91.8 89.9 87.5  
#   3310 2912 2556 2263 2032 1793 1548 1390 1246 1096 933 755 621 472 336 216 141 111 89 63  

Product Surveillance Registry Results

Number of Leads Enrolled in Study 4600
Cumulative Months of Follow-Up 308305
Number of Leads Active in Study 379
Information 105

Qualifying Complications

Cardiac Perforation 1
Conductor Fracture 42
Failure to Capture 8
Failure to Sense 2
Impedance Out of Range 13
Insulation (not further defined) 6
Lead Dislodgement 5
Oversensing 21
Other 4
Unspecified Clinical Failure 3
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 29
Conductor Fracture 26
Extra Cardiac Stimulation 2
Failure to Capture 83
Failure to Sense 36
Impedance Out of Range 61
Insulation Breach 4
Lead Dislodgement 124
Oversensing 142
Unspecified Clinical Failure 20
Information

Distribution Data

US Market Release2001-11-12
CE Approval Date2001-10-04
Registered USA Implants375859
Estimated Active USA Implants123350
US Market Release2001-11-12
CE Approval Date2001-10-04
Estimated WW Distribution619243
Information

USA Returned Product Analysis

Conductor Fracture 1451
Insulation Breach 104
Crimp/Weld/Bond 4
Other 199
Information

WW Returned Product Analysis

Conductor Fracture 1787
Insulation Breach 136
Crimp/Weld/Bond 4
Other 241
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar/Two Coils
More
Serial Number Prefix TDG
Lead Body Diameter 8.6 FR
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of December 1, 2025

www.medtronic.com/productperformance 22021