Product Education Brief: Micra Leadless Pacemaker - How to Prepare for Implant
Micraâ„¢ AV2/VR2 TPS devices
Original Date of Communication:
November 2025
View specific models that this applies to
Overview
This Product Education Brief re-enforces the instructions to program the real time clock (step 3 of the IFU below) along with completing other preparations before implanting a Micra device.
Applicable Product
Micra™ VR2 (MC2VR01), Micra™ AV2 (MC2AVR1)
Background on Micra device clock
The Micra devices initiate a date and time stamp at implant in order to trend data collected from the patient starting with the implant. If the internal Micra clock is not set before implant to match the geographic date and time of the implant location, data collection will proceed, but the associated time stamps on the data may not be as expected.
Preparing for Implant
The Micra instructions for use (IFU) describes the process of preparing a Micra device for implant prior to opening the sterile package and the order of operation should be followed. The Micra AV2/VR2 preparation steps are shown below.

If you have questions regarding this education brief, please contact Medtronic Technical Services.