Product Education Brief: Micra Leadless Pacemaker - How to Prepare for Implant

Product Education Brief: Micra Leadless Pacemaker - How to Prepare for Implant

Micraâ„¢ AV2/VR2 TPS devices
Original Date of Communication: November 2025

View specific models that this applies to

Overview

This Product Education Brief re-enforces the instructions to program the real time clock (step 3 of the IFU below) along with completing other preparations before implanting a Micra device.

Applicable Product

Micra™ VR2 (MC2VR01), Micra™ AV2 (MC2AVR1)

Background on Micra device clock

The Micra devices initiate a date and time stamp at implant in order to trend data collected from the patient starting with the implant. If the internal Micra clock is not set before implant to match the geographic date and time of the implant location, data collection will proceed, but the associated time stamps on the data may not be as expected.

Preparing for Implant

The Micra instructions for use (IFU) describes the process of preparing a Micra device for implant prior to opening the sterile package and the order of operation should be followed. The Micra AV2/VR2 preparation steps are shown below.

If you have questions regarding this education brief, please contact Medtronic Technical Services.


Specific Models This Applies To